• No results found

Influence of coaxial cable on response of high-voltage resisitive dividers

N/A
N/A
Protected

Academic year: 2021

Share "Influence of coaxial cable on response of high-voltage resisitive dividers"

Copied!
5
0
0

Loading.... (view fulltext now)

Full text

(1)

The 20th International Symposium on High Voltage Engineering, Buenos Aires, Argentina, August 27 – September 01, 2017

INFLUENCE OF COAXIAL CABLE ON RESPONSE OF

HIGH-VOLTAGE RESISTIVE DIVIDERS

A. Bergman1*, M. Nordlund1, A-P Elg1, J. Havunen2, J. Hällström2 and J. Meisner3 1

SP Sveriges Tekniska Forskningsinstitut AB, Box 857, 501 15 Borås, Sweden 2

VTT Technical Research Centre of Finland Ltd, Centre for Metrology MIKES Box 1000, 02044 VTT, Finland

3

PTB Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany

*Email: anders.bergman@sp.se

Abstract: An effort is pursued by several European National Measurement Institutes to

lower the uncertainties in calibration of UHV measuring systems for lightning impulse. To this end, several reference dividers are investigated as regards their accuracy both for amplitude and for time parameters. During these investigations a deterioration of step response was identified when longer coaxial cables were inserted in the measuring circuit. The measured front time T1 was also affected, in one observed case by 2.5 % elongation of front time as another 25 m cable was inserted. Compared to the intention to calibrate front time measurement to better than 5 % uncertainty for front time, this contribution must be well known, or preferably be eliminated. This paper presents the experimental findings from these investigations.

The investigated cables included selected coaxial, tri-axial, and cables with a corrugated screen. The effect of cable length was also studied. The influence was first discovered when applying a very fast step (rise-time < 4 ns) to the high voltage arm of a resistive divider and convolution of this step with the time derivative of an ideal lightning impulse with 0.84/60 µs impulse. The calculated output was analysed with IEC 61083 compliant software to evaluate the front time. Subsequently, these analyses have been augmented by additional comparative measurements where two reference dividers were connected to the same impulse generator, and varying the cable length of one of them. The summarized changes in front time calculated for different combinations of cable and impulse voltage dividers are shown and discussed. It is noted that a change in T1 error depends both on length of cable and its type. The results show that non-negligible front time errors may be introduced when the cable length is increased. To support these findings, further tests have been carried out with two reference impulse dividers connected in standard calibration configuration in accordance with IEC 60060-2. One divider was used as reference, while the cable for the other was varied. In this way, the change of error between configurations could be measured.

A theoretical study has also been performed, calculating the distortion of a lightning impulse on a coaxial cable. The results agree qualitatively with experiments, but the detailed results show discrepancies that need further investigation.

1 INTRODUCTION

An effort is pursued by several European National Measurement Institutes to lower the uncertainties in calibration of UHV measuring systems for lightning impulse. To this end, several reference dividers are investigated as regards their accuracy both for amplitude and for time parameters. During these investigations a deterioration of step response was identified when longer coaxial cables were inserted in the measuring circuit. The measured impulse parameters were all affected, amplitude scale factor, front time T1 and tail time T2. The effect could be up to several percent, which is not acceptable for reference measuring systems. A literature study has revealed that the phenomenon has been known since a long time, the earliest reference found is from 1932 [1, 2]! A

later study by Sato [3] has made a theoretical analysis of the effect, using the telegraph equation. The study shows that the effective resistance of the cable plays a central role.

2 EXPERIMENTAL INVESTIGATION 2.1 Method

The influence of the coaxial cable on the front time was primarily investigated by comparative measurements between reference measuring systems where the cable length was varied by inserting extra lengths of cable.

As a supplementary check, a very fast step (rise-time < 4 ns) was applied to the high voltage arm of a resistive divider. The step applied to the divider is generated by a mercury wetted relay based step

(2)

generator with an output voltage of 200 V. Evaluation of the step response is performed by convolution in accordance with IEC60060-2:2010. The step is convolved with the time derivative of an ideal lightning impulse with 0.84/60 µs impulse (front/tail time). The calculated output is analysed with IEC 61083 compliant software to evaluate the front time. The front time of the convolved curve is compared with the front time of the ideal lightning impulse where the difference gives a measure of the error introduced by the cable.

For one set of cables these analyses have been

augmented by additional comparative

measurements where two reference dividers were connected to the same impulse generator, and varying the cable length and type of one of them. Impulses of 400 kV with a front time of 0.84 µs were used during the high voltage comparison. A set of triaxial cables have been investigated with the same method as described above, convolving a step response with an ideal curve. In addition a lightning impulse generator, capable of producing 0.84/60 µs impulses with an amplitude of 2 kV, has been used as a supplementary method to investigate the front time error for different length of cables.

Both coaxial and triaxial cables have been tested to clarify if the influence on high voltage impulse measurements is related to a specific cable or not. Table 1. Cables investigated

Name Cable type Z0

[Ω] C [pF/m] R [mΩ /m]

Heliax ½” Coaxial Heliax 50±1 75.8 1.57

RG214 RG214 Coaxial 50 101 5.6 Sucofeed 3/8” Coaxial Sucofeed 3/8” high-flex 50±1 79.5 4.23 Triaxial* Triaxial 11 Camera Cable 75 54 14 Hyperflex * Triaxial 11mm Hyperflex 2000 75±2.2 54 13 Nokia* Triaxial 75 75 11.5 Belden 9888 Triaxial 50 85.3 3.94

*Same type of cables but different manufacturers.

Heliax ½” cable is intended for applications up to 3 GHz and exhibits 4 mm solid inner conductor (copper-clad aluminium) and corrugated copper screen, providing excellent high frequency characteristics.

RG214 is a ca. 10 mm diameter coaxial cable equipped with double screens for optimum screening efficiency and intended for applications up to 1 GHz.

The Sucofeed 3/8” cable is intended for applications up to 3 GHz and exhibits 2.8 mm solid inner conductor (copper-clad aluminium) and corrugated copper screen, providing excellent high frequency characteristics.

2.2 High voltage comparison

2.2.1 Setup

Two experiments were performed where the main difference was the length of the measuring cable for the “reference” measurement, i.e. the system that was unchanged while the other was equipped with different cable lengths. In one case the “reference” cable was 25 m and connecting to the recording device without use of shielded cabinet. In the other case, the “reference” cable was only 5 m and both cable screens were solidly connected to a high-quality shielded enclosure for the recording device. These two measurements give different results, leading to a surmise that the parallel connection of the cable shields affect the measurement. The same impulse voltage and waveshape were applied for each series of measurements.

2.2.2 15 m reference cable

The test was conducted with a waveshape of 0.84/41 µs at -100 kV. The “reference” was VTT MIKES 400 kV reference divider and their reference recording device, a NI5124 digitizer and interconnected with 15 m Belden 9888 triaxial cable. The other branch was a SMR700 reference divider equipped with successively longer RG214 cables.

Figure 1. Errors observed when changing length of signal cable. The reference cable is 15 m long.

While 75 m cables will be excessively long for normal calibration work, it enables clear results to be obtained, which can be used to evaluate impact also of shorter cables.

2.2.3 5 m reference cable

The test was conducted with two waveshapes, one of 0.84/41 µs and one at 1.56/45 µs both at -200 kV. The reference was SP-RISE 800 kV reference divider and their reference recording device, NI5124 digitizer and interconnected with 5 m triaxaxial cable. The other branch was a SMR700 reference divider equipped with

-1% 0% 1% 2% 3% 4% 0 20 40 60 80 0.84/41 µs Scale factor T1 error [%] T2 error [%]

(3)

successively longer RG214 cables. Some other cables were also tested.

Figure 2. Errors observed when changing length of signal cable. The reference cable is 5 m long.

Figure 3 Errors observed when changing length of signal cable. Reference cable is 5 m long.

2.2.4 Discussion

The difference between Figure 1 and Figure 2 is striking in that the induced changes of the errors are large, although, at first glance, they should be comparable. The main difference is however the arrangement of reference cable, which in the case of Figure 2 may act as a low-impedance path in parallel to the resistance of the shield of the “test” side. It is quite plausible that this will have an effect that at least qualitatively explains the observed difference.

2.2.5 Tests with other cables

It was deemed to be of interest to investigate the performance of high-quality GHz range cables. In this case two cables were available for these tests, listed above as Heliax and Sucofeed. Both are with solid center conductor embedded in foam dielectric. The shield is solid, corrugated copper. The findings are summarised in table below

Table 2. Experiment with radio-frequency cables Scale factor

error

T1 error T2 error

Heliax 1/2 -0.6 % -0.6 % -0.2 %

Sucofeed 3/8 0.0 % 0.3 % 0.4 %

Common feature of both cables is an appreciably lower DC resistance. It remains to be proven that this is the reason for the lower errors obtained with them. It should be noted that the zero-level of errors has not been firmly established, in fact, the authors believe that the results with Heliax is close to zero errors, although reported as negative errors.

In addition to these results, other cables such as the listed triaxial cables have been investigated. These cables have appreciably higher resistance and errors are also large. Here is presented only the front time error versus increased cable length, as measured with a low voltage impulse of 200 V, and theoretically calculated.

Figure 4. Triaxial cable induced errors for front time.

3 SIMULATED RESPONSE 3.1 Background

Although the transmission system, i.e. coaxial cable, should be included in every calibration of lightning impulse (LI) measuring systems, see [4], the importance of this requirement has often been neglected. In the course of measurements performed to characterise LI reference measuring systems, the authors observed distortions in the measured front-time, which could be traced to the use of different lengths of coaxial cables. A literature search revealed that this had been observed in 1932, in conjunction with design of delay lines for oscilloscopes [1, 2]. Later work by Sato [3] has discussed the phenomenon in relation to lightning impulse measurements, giving the theoretical background. His theory is summarised below.

3.2 Theory

The cable is assumed to be properly terminated at both ends, so that no reflections appear. The complex impedance of the cable is approximated by a real number to enable an analytical solution. -1% 0% 1% 2% 3% 4% 0 20 40 60 80 0.89/42 µs Scale factor T1 error T2 error -1% 0% 1% 2% 3% 4% 0 20 40 60 80 1.56/45 µs Scale factor T1 error T2 errror -1% 0% 1% 2% 3% 4% 0 10 20 30 40 50 60 0.84/60 µs 200V LI Step

(4)

Figure 5. Cable with impedance Z0 as a real number (from Sato [3]).

If the line constants are given as series resistance R (Ω/m), inductance L (H/m), leakage conductance G (S/m) and capacitance C (F/m), then the characteristic impedance z0 and propagation constant γ are defined in the frequency domain as:

R

s

L

 

G

s

C

C

s

G

L

s

R

z

,

0 (1)

Figure 6. Definition of coaxial cable parameters

Manipulating equation 1, it can be shown for perfect termination, relation between the voltages Es at the sending end and Er at the receiving end is

x s s r

E

e

x

x

E

E

)

sinh(

)

cosh(

1

(2)

Using the following variables to simplify

a

ca

b

RG

c

LG

RC

LC

a

2

4

d

,

,

b

,

2

(3)

For

E

s specified as a superposition of two exponential functions,

g

(

t

)

e

t

e

t, we apply the convolution integral and for t > t0 we obtain:

    d t t d I e t g t d e t t g E a b t t t a b r 2 0 2 2 0 2 1 2 0 2 0) ( ) ( 0 0        

 (4)

Where

I

1 is a modified Bessel function of the first kind, and

t

0

ax

is the surge propagation time along the cable.

This integral is not suitable for analytical solution, but numerical methods are readily applicable.

3.3 Simulation

The solution of equation 4 has been implemented in software and used to investigate the cables used in the experimental work. The capacitance of the cable has been taken as the nominal value, whereas the series resistance value has been varied to simulate on one hand the result of assuming that the DC resistance is valid and on the other hand taking into account the skin effect on effective conductor resistance. For the purpose of this simulation, skin effect has been calculated for the inner conductor based on its diameter and at a frequency of 1 MHz. The frequency has been chosen somewhat arbitrarily, but should be a reasonable estimate for work with lightning impulse.

Table 3. Calculated errors for different cables

Name Cable length Scale factor error T1 error T2 error [m] [%] [%] [%] Triaxial 25 0 0.5 2.8 Triaxial 50 -1.0 1.0 4.1 Triaxial 75 -2.1 1.6 5.4 RG214 25 0.2 0.3 2.6 RG214 50 -0.8 0.6 3.7 RG214 75 -1.7 1.0 4.8 Heliax 20 0.4 0.3 2.4 Sucofeed 30 0 0.4 2.9

As seen from Table 3, results are not a very good approximation for the observed experimental results. Especially tail time

T

2 is far from experiments. The reason for this is not apparent and more research is needed to explain this phenomenon.

Qualitatively, theory agrees with the experiments and lends some credibility to the surmise that the series resistance of the coaxial cable plays an important role

4 CONCLUSIONS

The phenomenon that the coaxial cable length will impact on the performance of resistive reference voltage dividers for lightning impulse has been “rediscovered”. The effect can have a large effect on the achievable performance of reference measuring systems. The possible impact on international standards [4] should be carefully evaluated.

(5)

ACKNOWLEDGMENTS

The work reported here has received support from the EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme.

REFERENCES

[1] F. P. Burch, "LXXI. On potential dividers for cathode ray oscillographs," The London, Edinburgh, and Dublin Philosophical Magazine and Journal of Science, vol. 13, pp. 760-774, 1932/04/01 1932.

[2] G. W. Bowdler, Measurements in high-voltage test circuits. Oxford: Pergamon, 1973.

[3] S. Sato, T. Yamaguchi, S. Nishimura, and S. Nishimura, "Influence of measuring cable on lightning impulse parameters," Electronics and Communications in Japan, vol. 93, pp. 1-7, 2010.

[4] IEC 60060-2: 2010, High-Voltage Test Techniques - Part 2: Measuring systems.

References

Related documents

In one splay chamber all the strands from one side span cable has been analysed in order to compare the sum of the axial forces in the 85 strands with the axial force in the

In this paper, the stator slot design of a cable wound permanent magnet synchronous generator for hydrokinetic energy conversion is studied using finite element (FE) simulations..

population, and also find no support for a suicide-provoking effect of these substances, but, in contrast, reduced rating of suicidal ideation in SSRI-treated

Briefly, these experiments indicate that more &#34;anxious&#34; rats display a gene expression profile suggesting a higher capacity for serotonin production, and are more prone

The specific aims of this thesis were: (i) to investigate the possible influence of serotonin-related genetic variation on the neural correlates of anxiety, and on mood-

To verify the propagation delay and reflection coefficient for lossless cables and attenuation for lossy cables, we introduced some transmission line theories

[…] Jag hoppas du trollar, för det där var sällsynt motbjudande (Flashback 2009). Inlägget ovan är också en del i konstruktionen som vi kan se. Köp av sexuella

In the capacity range up to 5,000 passengers per hour and over a track length of up to 5 kilometers, the CABLE Liner ensures maximum mobility in the urban environment. As well