• No results found

Using Deep Learning to Emulate the Use of an External Contrast Agent in Cardiovascular 4D Flow MRI

N/A
N/A
Protected

Academic year: 2021

Share "Using Deep Learning to Emulate the Use of an External Contrast Agent in Cardiovascular 4D Flow MRI"

Copied!
10
0
0

Loading.... (view fulltext now)

Full text

(1)

Using Deep Learning to Emulate the Use

of an External Contrast Agent in

Cardiovascular 4D Flow MRI

Mariana Bustamante, PhD,

1,2

*

Federica Viola, MSc,

1

Carl-Johan Carlhäll, MD, PhD,

1,2,3

and Tino Ebbers, PhD

1,2

Background: Although contrast agents would be beneficial, they are seldom used in four-dimensional (4D) flow magnetic resonance imaging (MRI) due to potential side effects and contraindications.

Purpose: To develop and evaluate a deep learning architecture to generate high blood–tissue contrast in noncontrast 4D flow MRI by emulating the use of an external contrast agent.

Study Type: Retrospective.

Subjects: Of 222 data sets, 141 were used for neural network (NN) training (69 with and 72 without contrast agent). Evalu-ation was performed on the remaining 81 noncontrast data sets.

Field Strength/Sequences: Gradient echo or echo-planar 4Dflow MRI at 1.5 T and 3 T.

Assessment: A cyclic generative adversarial NN was trained to perform image translation between noncontrast and con-trast data. Evaluation was performed quantitatively using concon-trast-to-noise ratio (CNR), signal-to-noise ratio (SNR), struc-tural similarity index (SSIM), mean squared error (MSE) of edges, and Dice coefficient of segmentations. Three observers performed a qualitative assessment of blood–tissue contrast, noise, presence of artifacts, and image structure visualization. Statistical Tests: The Wilcoxon rank-sum test evaluated statistical significance. Kendall’s concordance coefficient assessed interobserver agreement.

Results: Contrast in the regions of interest (ROIs) in the NN enhanced images increased by 88%, CNR increased by 63%, and SNR improved by 48% (allP < 0.001). The SSIM was 0.82 ± 0.01, and the MSE of edges was 0.09 ± 0.01 (range [0,1]). Segmentations based on the generated images resulted in a Dice similarity increase of 15.25%. The observers managed to differentiate between contrast MR images and our results; however, they preferred the NN enhanced images in 76.7% of cases. This percentage increased to 93.3% for phase-contrast MR angiograms created from the NN enhanced data. Visual grading scores were blood–tissue contrast = 4.30 ± 0.74, noise = 3.12 ± 0.98, and presence of artifacts = 3.63 ± 0.76. Image structures within and without the ROIs resulted in scores of 3.42 ± 0.59 and 3.07 ± 0.71, respectively (P < 0.001). Data Conclusion: The proposed approach improves blood–tissue contrast in MR images and could be used to improve data quality, visualization, and postprocessing of cardiovascular 4Dflow data.

Evidence Level: 3

Technical Efficacy: Stage 1

J. MAGN. RESON. IMAGING 2021.

I

n conventional magnetic resonance (MR) angiography, con-trast agents are crucial to generate blood–tissue concon-trast.1

However, their use may be contraindicated in patients with renal impairment,2,3 and recent reports have shown some deposition in the brain after the examination.4,5

Phase-contrast magnetic resonance imaging (PC-MRI) acquisition techniques can be employed to generate angio-graphic images without the use of contrast agents. Among them, four-dimensional (4D) flow MRI can be obtained by using flow encoding in three spatial directions to generate

View this article online at wileyonlinelibrary.com. DOI: 10.1002/jmri.27578

Received Nov 12, 2020, Accepted for publication Feb 13, 2021.

*Address reprint requests to: M.B., 581 85 Linköping, Sweden. E-mail: mariana.bustamante@liu.se

From the1Division of Diagnostics and Specialist Medicine, Department of Health Medicine and Caring Sciences, Linköping University, Linköping, Sweden; 2Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden; and3Department of Clinical Physiology in Linköping

and Department of Health Medicine and Caring Sciences, Linköping University, Linköping, Sweden Additional supporting information may be found in the online version of this article

This is an open access article under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non-commercial and no modifications or adaptations are made.

(2)

time-resolved volumes containing both morphology and blood-flow velocity information.6,7

These content-heavy images require longer acquisition times and typically result in lower blood–tissue contrast when compared with routine car-diac MRI. Consequently, their focus is commonly not on morphological diagnostics but on functional assessment of the blood flows throughout the cardiovascular system. These images would benefit from techniques that improve their quality at the postprocessing stage.

Using deep learning, the task of generating an image of a specific class has been solved by training a neural network (NN) to infer the probability distribution of the input data and generate new samples in this distribution.8 Generative adversarial networks (GANs) are a type of NN in which two entities—a generator and a discriminator—are trained simul-taneously using ideas originated from game theory.9The gen-erator focuses on understanding the distribution of the data, while the discriminator estimates the probability of a sample belonging to the true distribution rather than having been created by the generator. This approach has been used to gen-erate highly realistic images in several categories, including medical imaging.8,10Recently, Zhu et al11 achieved excellent performance in image-to-image translation using cycle-consistent adversarial networks (CycleGANs), a variation of the GAN model that aims to translate images from a specific domain A to domain B, and vice versa. A CycleGAN can be treated as unsupervised learning,10 only requiring a set of images for each domain, without the need for pairs of images corresponding to each input. This is a significant advantage in medical imaging, where acquiring multiple data sets from each subject is typically unfeasible. The CycleGAN has been successfully used for synthesis and denoising of medical images.12,13

Thus, the aim of this study was to develop and evaluate a CycleGAN architecture to emulate the effect of a contrast agent in 4Dflow MR images acquired without using contrast agents.

Materials and Methods

The research was performed in line with the Declaration of Helsinki and was approved by the regional ethics board. The exams were per-formed specifically for research purposes. All subjects gave written informed consent.

Study Data

A total of 222 4Dflow MR data sets were used in this study. From this total, 141 were included in the training process: 72 were acquired without contrast medium, and 69 directly after a gadolin-ium contrast agent (Magnevist; Bayer Schering Pharma AG) was injected into the subjects prior to the acquisition of a late-enhancement study. An additional group of 81 4Dflow MRI data sets acquired without contrast medium was used to test the trained network and evaluate the results. The training set of 141 subjects included 59 healthy volunteers, 27 with and 32 without contrast. There were 46 healthy volunteers in the test set. The included

patients represented a wide range of medical disorders including chronic ischemic heart disease, idiopathic dilated cardiomyopathy, diastolic heart failure, history of atrialfibrillation, mild to moderate mitral valve regurgitation, postmitral valve repair, and diabetes type 2. The study data did not include patients with congenital heart dis-ease. Exclusion criteria for both groups (contrast and noncontrast) were contraindication for MRI and significantly irregular ventricular rhythm. Exclusion criteria for the contrast group were very low qual-ity due to the presence of artifacts and/or noise and timing issues during the acquisition generating very low blood–tissue contrast in the magnitude images.

Free-breathing, respiratory-motion-compensated, 4D flow examinations were acquired on 1.5-T and 3-T MRI scanners (Ingenia; Philips Healthcare, The Netherlands). Scan parameters included sagittal-oblique slab covering the whole heart and thoracic aorta, velocity encoding (VENC) 120 cm/sec–150 cm/sec, flip angle 5–10, echo time 2.5 msec–5.0 msec, repetition time 4.2 msec– 9.1 msec, Sensitivity Encoding (SENSE) speed up factor 3 (AP direction) or 4 (2 in Anterior-Posterior (AP) and 2 in the Right-Left (RL) direction), k-space segmentation factor 2–3 for gradient-echo and read-out factor 3–7 for echo-planar-imaging-based sequence, acquired temporal resolution of 30.0 msec–52.8 msec, and spatial resolution 3 mm3. Typical scan time was 4–10 minutes, with respiratory navigator efficiency of 60%–80%.

Data Preprocessing

Each 4D flow MR magnitude volume was sliced in the anteroposterior direction to obtain images of the cardiovascular sys-tem’s coronal plane. Each image was then resampled to the same size of 128× 128. A segmentation of the heart and major vessels gener-ated using an automatic atlas-based method was used to locate the main regions of interest (ROIs).14 A rectangle encompassing these regions was used to exclude areas of large dissimilarity between dif-ferent subjects, such as the chest, abdomen, and spine. Finally, two groups (contrast and noncontrast) of 117,120 two-dimensional (2D) images each were used to train the network. Only the magni-tude image included in each 4Dflow MR data set was used as input to the network; consequently, the velocity information included in the acquisition was not altered in any way.

Network Architecture

The CycleGAN model was composed of two discriminators and two generators that achieve translation between two domains, A and B. In addition, the networks were trained to minimize cycle consis-tency losses,11 which aim to guarantee consistency when forward

and backward translations are applied successively on an image. A detailed depiction of the network can be seen in Fig. 1.

The model was implemented using TensorFlow15and trained for

approximately 2.5 epochs with the following parameters: batch size = 24, generator with six residual blocks,16 Adam optimizer,17 and learning

rate = 0.0002. Residual blocks were composed of two convolutional layers followed by instance normalization,18and Rectified Linear Unit

(ReLU) activation.

Training was performed on a workstation with a 3.6-GHz, six-core processor with 64-GB RAM, NVIDIA Quadro P6000 GPU.

(3)

Evaluation

Evaluation of the NN enhanced images was performed quantitatively and qualitatively using metrics that focus on the main areas of inter-est: blood–tissue contrast, noise, artifacts, and structural consistency.

The quantitative evaluation was performed on the original data, the NN enhanced data, and, for reference, data on which contrast-limited adaptive histogram equalization (CLAHE)19 was applied. The following metrics were used:

• Signal difference (contrast) between ROIs (heart and great vessels) and the remaining image.

• Contrast-to-noise ratio (CNR) calculated as

CNR =j SROI−SBgj σo

where SROI and SBg are the signal intensities in the ROI and the

remaining background, respectively; andσois the standard deviation

of the noise.

• Signal-to-noise ratio (SNR) in the whole image calculated as

SNR =μim σo whereμimis the average signal in the image.

• SNR in the ROIs calculated as

SNR =μROI σo

whereμROIis the average signal in the ROIs.

• The structural similarity index (SSMI), a quality assessment met-ric, used to compare an image to its reference focusing on struc-tural information while also incorporating luminance and contrast components.20The resulting index is in the range [0, 1], where 1 indicates that the two images are identical.

• Mean squared error (MSE) of the outlines generated using Canny edge detection21between the original images and NN enhanced images.

Additional quantitative evaluation was performed on the utility of the NN enhanced images to create three-dimensional (3D) segmentations of the heart and great vessels. Phase-contrast MR angiograms (PC-MRAs) were created from the NN enhanced as well as from the original data by combining the magnitude and velocity information contained in the 4Dflow MRI, as proposed in the study by Bock et al.22 An appropriate threshold for each PC-MRA was found using Otsu’s method.23 Segmentations of the

car-diac chambers, aorta, and pulmonary artery were also generated automatically from the original data using an atlas-based method,14 which after manual correction served as a reference standard. Com-parison of the PC-MRA segmentations with reference standard was performed using Dice similarity coefficient.24

Qualitative evaluation of the results was performed by three observers (one imaging expert and two clinicians: T.E., C-J.C., and J.S.) with 22, 16, and 4 years’ experience with 4D flow MRI, respec-tively. The evaluation was done using a questionnaire with the fol-lowing tasks:

• Task 1: The observer received 30 images and was asked whether each image was a contrast MR image or an NN enhanced image.

FIGURE 1: Network architecture. Top: General design of the training process. GABis the forward generator that transforms and

element “a” from group A to group B, resulting in “b’”; and vice versa for the backward generator GAB. DA and DB are the

discriminators in A and B domains, respectively. Cycle consistency losses are calculated between an original element and the result of applying the two generators consecutively. Bottom: Generator and discriminator architectures. Each layer includes the number of filters, filter size, and stride.

(4)

• Task 2: The observer received 20 image pairs consisting of an original (noncontrast) 4Dflow image and its processed equivalent. The observer then graded blood–tissue contrast, noise level, and artifact level of the NN enhanced image in comparison to the original using the scales (1–5) provided in Table 1. Comparative image structure visualization was graded using the scale (1–4) pro-vided in Table 2.

• Task 3: The observer received 20 blinded image pairs consisting of an original (noncontrast) 4D flow image and its processed equivalent and was asked which one they believed had better quality.

• Task 4: The observer received 10 blinded video pairs of PC-MRAs generated from an original (noncontrast) 4D flow MR image and its processed equivalent and was asked which one they believed had better quality.

The tasks were done without washout in the sequence order previously described. Each reader performed their analysis independently.

Statistical Analysis

All differences between quantitative parameters (noncontrast original data vs. NN enhanced) were evaluated for statistical significance using the Wilcoxon rank-sum test. A P-value of 0.001 was consid-ered statistically significant.

Kendall’s concordance coefficient (W) was used to describe the interobserver agreement of qualitative parameters.25

Results

The CycleGAN model was successfully created and trained, as demonstrated by the examples from the original non-contrast MR images and their corresponding NN enhanced images shown in Fig. 2. A comparison of PC-MRAs gener-ated from 4D flow MRI before and after applying the pro-posed method can be seen in Fig. 3. Only 2D images are included in this section due to space constraints, and 3D ver-sions of these images in video format have been included as in the Supplementary Material.

Quantitative Evaluation

Figure 4(a,b) shows the comparison of signal difference (con-trast) and CNR between the original images, the NN enhanced images, and images after application of adaptive histogram equal-ization. Quantitative evaluation showed that the NN enhanced images had significantly higher contrast and CNR in the heart and great vessels (both P < 0.001): 0.47 ± 0.05 vs. 0.25 ± 0.04, and 18.0 ± 2.58 vs. 11.03 ± 2.59 (an increase in contrast of 88% and an increase in CNR of 63%). Application of adaptive histogram equalization produced higher contrast when compared with the original images (0.32 ± 0.05); however, CNR decreased to 6.04 ± 1.09. These comparisons were also statistically signi fi-cant (P < 0.001).

A similar comparison was done for the SNR computed in the whole image and in the heart and great vessels and can be seen in Fig. 4(c,d). The effect of the proposed method was statistically significant (P < 0.001) for increasing SNRs in both cases, especially in the heart and great vessels: 14.49 ± 1.9 vs. 17.63 ± 1.65, and 18.01 ± 2.65 vs. 26.6 ± 2.32 (an increase of 21% for the whole image, and 48% for the heart and great vessels). Adaptive histogram equalization resulted in significantly lower SNR values on both cases (P < 0.001): 9.61 ± 0.69 for the whole image, and 11.0 ± 0.92 for the heart and great vessels.

TABLE 1. Image Quality

Grades Blood–tissue contrast Noise Artifacts 1 Significantly less contrast than

original

Significantly more noise than original

Significantly more artifacts than original

2 Slightly less contrast than original Slightly more noise than original Slightly more artifacts than original 3 Identical or nearly identical to

original

Identical or nearly identical to original

Identical or nearly identical to original

4 Slightly more contrast than original Slightly less noise than original Slightly less artifacts than original 5 Significantly more contrast than

original

Significantly less noise than original

Significantly less artifacts than original

TABLE 2. Image Structure

Grades

Structural Changes Within the Heart and Major Vessels

Structural Changes Outside the Heart and Major Vessels 1 Evident changes visible Evident changes visible 2 Moderate changes visible Moderate changes visible

3 Slight changes visible Slight changes visible 4 No changes visible No changes visible

(5)

The metrics related to structural consistency (Fig. 5) also resulted in positive outcomes, with the SSIM having an average close to the maximum value of one (0.82 ± 0.01), while the MSE of the com-parison of edges between the images was quite low (0.09 ± 0.01).

With respect to the segmentation component of the evaluation, Dice similarity coefficients of the segmentations created using the PC-MRAs of the original images and the NN enhanced ones can be seen in Fig. 6. The average Dice coefficient for the original images was 0.59 ± 0.1, while the

average for the NN results was 0.68 ± 0.06, representing an increase of 15.25% (P < 0.001).

Qualitative Evaluation

A summary of the results obtained during the qualitative evalua-tion is shown in Fig. 7. It was relatively easy for the observers to differentiate between real contrast MR images and the NN enhanced images; as can be seen in the results for task 1, where the observers were able to identify the correct option in 84.4% of the cases. However, when the choice was between an original noncontrasted MR image and an NN enhanced one (tasks

FIGURE 2: Example results. Top: Original magnetic resonance (MR) images acquired without contrast. Middle: Result of applying the proposed method. Bottom: Result of applying contrast-limited adaptive histogram equalization (CLAHE) to the original images.

FIGURE 3: Example phase-contrast magnetic resonance (MR) angiography results. Top: Phase-contrast MR angiograms (PC-MRAs) generated from 4Dflow MR images acquired without contrast. Bottom: PC-MRAs generated after applying the proposed method.

(6)

FIGURE 4: Quality measures. Blue: original magnetic resonance (MR) images, red: neural network (NN) enhanced images, and orange: images after application of contrast-limited adaptive histogram equalization (CLAHE). (a) Contrast difference between the region of interest (heart and great vessels) and the remaining image. (b) Contrast-to-noise ratio (CNR) at the region of interest (heart and great vessels) compared with the remaining image. (c) Signal-to-noise ratio (SNR) in the whole image. (d). SNR in the heart and great vessels. Mean and standard deviation for each group is also indicated.

FIGURE 5: Structural measures. (a) Structural similarity index (SSIM) of the neural network (NN) enhanced images using the original magnetic resonance (MR) images as reference. (b) Mean squared error (MSE) of the edges in the original MR images vs. NN enhanced images. Mean and standard deviation for each group is also indicated.

(7)

3 and 4), the proposed method’s results were chosen more fre-quently, especially when comparing angiographic images (76.7% in the case of the MR images and 93.3% for the PC-MRAs).

Detailed qualitative scores assigned by the observers dur-ing task 2 are shown in Table 3, and bar plots of the percent-ages obtained for each score in the scale can be seen in Fig. 7. Regarding blood–tissue contrast, which was the main focus of the study, the observers rated the NN enhanced images with an average of 4.30 ± 0.74 on a scale from 1 to 5, which repre-sents a slight to considerable improvement when compared with the original MR images acquired without contrast. Quali-tative evaluation of noise resulted in a mean of 3.12 ± 0.98, which (from Table 1) can be interpreted as no evident differ-ences detected between the original and NN enhanced images on average. Visual evaluation of the presence of artifacts in the images before and after the application of the proposed method resulted in a mean of 3.63 ± 0.76, which (from Table 1) suggested a moderate improvement in this category.

With respect to the image structure, the average score obtained when focusing on the heart and great vessels was 3.42 ± 0.59, and the average in the rest of the image was 3.07 ± 0.71. On the suggested scale of 1–4 (Table 2), this corresponds to slight or no changes visible in the ROIs, while

FIGURE 6: Comparison of segmentations generated using the phase-contrast magnetic resonance angiograms (PC-MRAs). Blue: PC-MRA computed using the original MR images, and red: PC-MRA computed using the neural network (NN) enhanced magnitude images. Mean and standard deviation for each group is also indicated.

FIGURE 7: Qualitative test results. Tasks 1,3, and 4 are depicted using pie charts to represent the observers’ answers. Task 2 is shown using bar plots of the percentages of scores assigned to the neural network (NN) enhanced images when compared to the original noncontrast magnetic resonance images.

(8)

slight changes were found to be present in the areas outside these regions.

Percentage agreement between the observers was high overall, ranging from 86% to 94%; Kendall’s concordance coefficient ranged from 0.54 to 0.63, also denoting substan-tial strength of agreement between raters. Detailed results can be seen in Table 3.

Discussion

In this study, we have developed and evaluated a deep learning-based method for the emulation of the effect of con-trast agent on nonconcon-trast cardiac 4D flow MRI. The tech-nique resulted in an overall increase in contrast and lower artifact visibility, especially in ROIs within the thoracic cav-ity. This increase in quality could improve the utility of 4D flow MRI in the clinical setting and may simplify the calcula-tion of funccalcula-tional parameters derived from these images.

Comparison with a classic technique such as adaptive histogram equalization served to demonstrate the superiority of the proposed method. Histogram equalization intensifies the artifacts and noise present in the original MR image and is only able to increase the general signal present in the image, without focusing on specific ROIs.

MRI data were acquired using two different MR scan-ners (1.5 T and 3 T). Although we did not explore the possi-ble distinctions quantitatively, visual inspection showed no major quality differences due to the different scanners used. However, because both scanners are used regularly in our clinical routine, it was important for the NN to train in a group of images as diverse as possible. This was also the rea-son behind the inclusion of patients with a wide variety of diseases in addition to healthy volunteers.

The proposed technique can be applied in approxi-mately 2 minutes for a full cardiac 4Dflow MRI composed of approximately 2000 2D images. This method may there-fore represent a fast way of improving image quality without extending the already long acquisition time of these images.

The generator network in the proposed model pro-duces images by applying a succession of operations to the original image. The first section “encodes” the input image into a set of features of different sizes, while the second “decodes” these features into a result. It is therefore possible that the generated images could be affected by structural changes unintentionally added by this process, especially if the network suffers from overfitting.26However, the quanti-tative metrics related to structure similarity obtained (SSIM and MSE of the edges) showed good agreement with the original images, and the qualitative scoring resulted in no obvious differences. It is important to note that the SSIM can be hard to interpret in this situation because it typically expects the reference image (the original MRI in this case) to be of better quality than the evaluated image, which is not necessarily true here. In addition, during the qualitative test, the observers were required to focus on either the main ROIs (heart and great vessels) or on the remaining sections outside this area. In this case, the goal was to obtain an eval-uation specific to the areas corresponding to the heart and vessels because structural differences in these regions would negatively affect the results the most. The qualitative results obtained suggest that structural changes were indeed less vis-ible within the heart and vessels, if found at all. Qualitative evaluation also indicated a moderate decrease in both noise and artifact visualization in the NN enhanced images. This was probably due to the higher overall quality of the contra-sted training group.

The improvement observed with regard to the PC-MRAs could be particularly useful during postprocessing and analysis because these images are heavily used to guide vessel segmentation and visualization in 4Dflow MRI.22,27 Genera-tion of a 3D segmentaGenera-tion derived from the NN enhanced angiograms resulted in higher Dice similarity indexes than segmentations generated using PC-MRAs from original non-contrast data when compared to reference standard segmenta-tions of the heart, aorta, and pulmonary veins. It is worth noting that the Dice indexes in this test were not expected to

TABLE 3. Qualitative Scores Obtained for Each Metric, Percentage of Agreement Between Observers, and Kendall’s Concordance Coefficient (W)

Metric 1 2 3 4 5 Total % Agreement W Blood–tissue contrast 0 0 4 30 26 60 94% 0.58 Noise 0 17 27 8 8 60 86% 0.61 Artifacts 0 3 23 27 7 60 93% 0.54 Metric 1 2 3 4 Total % Agreement W Structure (Heart and vessels) 0 3 29 28 60 92% 0.54 Structure (Rest of image) 1 10 33 16 60 86% 0.63

(9)

be close to 1.0 because the reference standard segmentations included the main cardiac and vessel regions (cardiac cham-bers, aorta, and pulmonary artery) but excluded vessels such as the venae cavae, pulmonary veins, carotid arteries, and distal branches of the pulmonary artery.

For the purposes of training the model, areas of large dissimilarity such as the subject’s chest, abdomen, and spine were removed from the images. This made both contrast and noncontrast training groups more consistent and encouraged the network to focus on ROIs within the thoracic cavity.

One of the main advantages of the chosen network architecture is that it does not require paired training exam-ples. While it is possible that the results could be improved with the use of paired data, it is quite unlikely that a large number of paired data would be available since the use of contrast agents and the number of 4Dflow MRI acquisitions are usually restricted to the minimum necessary due to side effects and higher costs.

Although no unexpected differences in contrast between slices of the same data set were apparent in the results, a simi-lar network architecture to that currently used could be implemented to receive and generate 3D volumes including the entire ROI at one time frame. However, 3D GANs have not been explored as extensively as 2D ones, would require significantly more resources during training, and there is no guarantee that they would improve the presented results due to the added complexity.

Limitations

The contrast agent used in our training group was injected as a bolus for a late gadolinium enhancement study. The 4D flow MR images were acquired during the waiting time between the injection of contrast and the gadolinium enhancement acquisition. Consequently, blood–tissue con-trast was higher in the 4Dflow MR images when compared with images acquired without contrast, but not as high as if an intravascular contrast agent had been used. Training on 4Dflow MRI data acquired with intravascular contrast agents could have improved the results further. However, these con-trast agents are rarely used, and such data were unavailable.

Images corresponding to several different studies (using different sequences andfield strengths) had to be included in order to reach a sufficiently large group to use during train-ing. This can be viewed as an advantage for the training pro-cess because it resulted in a training set with higher variability. However, as a consequence of this, not all the images used were of the same width and height and had to be resized, potentially affecting the sharpness of a portion of the data used. It is possible that the presented results could be improved further by including only images of the same size.

A few subjects presenting medical conditions that could alter the flow dynamics, such as valvular regurgitation, were included. These issues can introduce some signal loss in the

magnitude images due to turbulence or jet flows possibly affecting the results; however, we did not study this speci fi-cally during the project since these data sets were a minority within our training set.

Finally, only magnitude images were used to train the presented model; however, it may be useful to include addi-tional information related to the velocity-encoded data. For instance, PC-MRA data could perhaps serve as a more defined ROI for the network to focus on during training.

Conclusion

Our results demonstrated the potential of applying a CycleGAN model to improve image quality in 4Dflow MRI, with special focus on increasing contrast in localized ROIs within the heart and great vessels. The implemented model only makes use of the magnitude component of the MR acquisition, having no impact on the velocity components. The proposed approach is fast and can be applied as a post-processing step, and the resulting images could potentially simplify the application of other postprocessing techniques routinely employed to assess cardiovascular function using these MR images.

Acknowledgments

The authors would like to acknowledge Jonathan Sundin (J. S.) for his valuable help during the qualitative evaluation of the results. The authors would also like to thank NVIDIA for donating the GPU used during this study. This work was funded by Sweden’s Innovation Agency Vinnova, project 2017-02447; the Swedish Research Council, grant number 2018-04454; the Swedish Medical Research Council, grant number 2018-02779; the Swedish Heart and Lung Founda-tion, grant numbers 20180657 and 20170440; and the Council of Östergötland, grant number LIO-797721.

References

1. Prince MR, Yucel EK, Kaufman JA, Harrison DC, Geller SC. Dynamic gadolinium-enhanced three-dimensional abdominal MR arteriography. J Magn Reson Imaging 1993;3(6):877-881.

2. Forghani R. Adverse effects of gadolinium-based contrast agents. Top Magn Reson Imaging 2016;25:163-169.

3. Behzadi AH, Zhao Y, Farooq Z, Prince MR. Immediate allergic reactions to gadolinium-based contrast agents: A systematic review and meta-analysis. Radiology 2018;286:471-482.

4. Kanda T, Fukusato T, Matsuda M, et al. Gadolinium-based contrast agent accumulates in the brain even in subjects without severe renal dysfunction: Evaluation of autopsy brain specimens with inductively coupled plasma mass spectroscopy. Radiology 2015;276(1):228-232.

5. Ramalho J, Ramalho M, Jay M, Burke LM, Semelka RC. Gadolinium tox-icity and treatment. Magn Reson Imaging 2016;34:1394-1398.

6. Markl M, Kilner PJ, Ebbers T. Comprehensive 4D velocity mapping of the heart and great vessels by cardiovascular magnetic resonance. J Cardiovasc Magn Reson 2011;13(1):7.

7. Stankovic Z, Allen BD, Garcia J, Jarvis KB, Markl M. 4Dflow imaging with MRI. Cardiovasc Diagn Ther 2014;4(2):173-192.

(10)

8. Yi X, Walia E, Babyn P. Generative adversarial network in medical imag-ing: A review. Med Image Anal 2019;58:101552.

9. Goodfellow I, Pouget-Abadie J, Mirza M, et al. Generative adversarial nets. Adv Neural Inf Process Syst 2014;27.

10. Creswell A, White T, Dumoulin V, Arulkumaran K, Sengupta B, Bharath AA. Generative adversarial networks: An overview. IEEE Signal Process Mag 2018;35:53-65.

11. Zhu JY, Park T, Isola P, Efros AA. Unpaired image-to-image translation using cycle-consistent adversarial networks. Proceedings of the IEEE international conference on computer vision; Venice, Italy: IEEE; 2017.

12. Wolterink JM, Dinkla AM, Savenije MHF, Seevinck PR, van den Berg CAT, Išgum I. Deep MR to CT synthesis using unpaired data. Lec-ture notes in computer science (including subseries lecLec-ture notes in artificial intelligence and lecture notes in bioinformatics); Cham, Switzerland: Springer International Publishing; 2017.

13. Kang E, Koo HJ, Yang DH, Seo JB, Ye JC. Cycle-consistent adversarial denoising network for multiphase coronary CT angiography. Med Phys 2019;46:550-562.

14. Bustamante M, Gupta V, Forsberg D, Carlhäll CJ, Engvall J, Ebbers T. Automated multi-atlas segmentation of cardiac 4D flow MRI. Med Image Anal 2018;49:128-140.

15. GoogleResearch. TensorFlow: Large-scale machine learning on hetero-geneous systems. 2015. https://arxiv.org/abs/1603.04467.

16. He K, Zhang X, Ren S, Sun J. Deep residual learning for image recogni-tion. 2016 IEEE conference on computer vision and pattern recognition (CVPR); Las Vegas, NV, USA: IEEE; 2016. p 770-778.

17. Kingma DP, Ba JLA. A method for stochastic optimization. 3rd interna-tional conference on learning representations, ICLR conference track proceedings; San Diego, CA, USA: ICLR; 2015.

18. Ulyanov D, Vedaldi A, Lempitsky V. Instance normalization: The missing ingredient for fast stylization. 2016. https://arxiv.org/abs/1607.08022.

19. Zuiderveld K. Contrast limited adaptive histogram equalization. Graphic Gems IV. USA: Academic Press Professional, Inc.; 1994; 474-485.

20. Wang Z, Bovik AC, Sheikh HR, Simoncelli EP. Image quality assess-ment: From error visibility to structural similarity. IEEE Trans Image Pro-cess 2004;13:600-612.

21. Canny J. A computational approach to edge detection. IEEE Trans Pat-tern Anal Mach Intell 1986;8:679-698.

22. Bock J, Wieben O, Johnson K, Hennig J, Markl M. Optimal processing to derive static PC-MRA from time-resolved 3D PC-MRI data. Proceed-ings 16th scientific meeting: Berlin, Germany: International Society for Magnetic Resonance in Medicine; 2007. p 3053.

23. Otsu N. A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern 1996;9(1):62-66.

24. Dice LR. Measures of the amount of ecologic association between spe-cies. Ecology 1945;26:297-302.

25. Gisev N, Bell JS, Chen TF. Interrater agreement and interrater reliabil-ity: Key concepts, approaches, and applications. Res Soc Adm Pharm 2013;9(3):330-338.

26. Cohen JP, Luck M, Honari S. Distribution matching losses can halluci-nate features in medical image translation. Medical Image Computing and Computer Assisted Intervention -- MICCAI 2018; Cham, Switzerland: Springer International Publishing; 2018.

27. Hennemuth A, Friman O, Schumann C, et al. In: Wong KH, Holmes IIIDR, editors. Fast interactive exploration of 4D MRIflow data. Medical Imaging 2011: Visualization, Image-Guided Procedures, and Modeling: 7964 USA: SPIE; 2011. p 110–120.

References

Related documents

After the last pooling layer the window size is 1x1 over 1000 channels, followed by a single fully connected layer of 1000 units, giving 1 million parameters after the

Patients with abnormal striatal uptake of DaTSCAN (Grades 1-4) were grouped together and were compared to patients with normal striatal uptake (Grade 5) in terms

Begränsas kommunikationen kring olika förändringar till stormöten och skriftlig information kan det vara svårare för chefer och andra medarbetare att förstå varför en

We validated our findings in two case studies and discussed three broad factors which were found to be critical for the formation of data collaboratives: value proposition, trust,

Hypotesen var delvis rätt då de yngre eleverna har en mer positiv inställning till föräldramedverkan i skolan, men vi har fel i hypotesen då det inte är någon större

Den första besvarar han med att det är de sex globaliseringsfaserna som fört oss hit och ur dessa kommer den inriktning som vi kallar modernismen, som spridit sig över plane- ten

1641, 2018 Department of Medical and Health Sciences. Linköping University SE-581 83

Studies using both synthetic and authentic data shows that the marginalized particle filter can in fact give better performance than the extended Kalman filter.. However,