• No results found

Magnetic Scanning Probe Calibration Using Graphene Hall Sensor

N/A
N/A
Protected

Academic year: 2021

Share "Magnetic Scanning Probe Calibration Using Graphene Hall Sensor"

Copied!
5
0
0

Loading.... (view fulltext now)

Full text

(1)

Magnetic Scanning Probe Calibration Using

Graphene Hall Sensor

Vishal Panchal, Oscar Iglesias-Freire, Arseniy Lartsev, Rositsa Yakimova, Agustina Asenjo

and Olga Kazakova

Linköping University Post Print

N.B.: When citing this work, cite the original article.

©2013 IEEE. Personal use of this material is permitted. However, permission to

reprint/republish this material for advertising or promotional purposes or for creating new

collective works for resale or redistribution to servers or lists, or to reuse any copyrighted

component of this work in other works must be obtained from the IEEE.

Vishal Panchal, Oscar Iglesias-Freire, Arseniy Lartsev, Rositsa Yakimova, Agustina Asenjo

and Olga Kazakova, Magnetic Scanning Probe Calibration Using Graphene Hall Sensor,

2013, IEEE transactions on magnetics, (49), 7, 3520-3523.

http://dx.doi.org/10.1109/TMAG.2013.2243127

Postprint available at: Linköping University Electronic Press

(2)

Magnetic scanning probe calibration using graphene Hall sensor

Vishal Panchal

1, 2

, Oscar Iglesias-Freire

3

, Arseniy Lartsev

4

, Rositza Yakimova

5

, Agustina Asenjo

3

and Olga Kazakova

1

1

National Physical Laboratory, Teddington, TW11 0LW, United Kingdom

2Royal Holloway, University of London, Egham, TW20 0EX, United Kingdom 3Instituto de Ciencia de Materiales de Madrid, CSIC, Cantoblanco, 28049, Madrid, Spain

4Chalmers University of Technology, Göteborg, S-412 96, Sweden 5

Linköping University, Linköping, S-581 83, Sweden

Magnetic force microscopy (MFM) offers a unique insight into the nanoscopic scale domain structures of magnetic materials. However, MFM is generally regarded as a qualitative technique and, therefore, requires meticulous calibration of the magnetic scanning probe stray field (Bprobe) for quantitative measurements. We present a straightforward

calibration of Bprobe using scanning gate microscopy on epitaxial graphene Hall sensor in conjunction with Kelvin probe

force microscopy feedback loop to eliminate sample-probe parasitic electric field interactions. Using this technique, we determined Bprobe ~70 mT and ~76 mT for probes with magnetic moment ~1×10

-13

and >3×10-13 emu, respectively, at a probe-sample distance of 20 nm.

Index Terms— Epitaxial graphene, Hall sensor, Kelvin probe force microscopy, magnetic probe calibration.

I. INTRODUCTION

AGNETIC force microscopy (MFM) is a well-established modification of atomic force microscope (AFM) technique for imaging of magnetic domains, allowing for effective mapping with nanoscopic spatial resolution. However, the technique is generally qualitative and further requires meticulous calibration of the stray magnetic field (Bprobe) of the scanning probe for calibrated quantitative measurements [1], [2]. Microscopic Hall sensors are ideal for such probe calibration [3–5], in particular graphene-based Hall sensors benefit from high sensitivity (Hall coefficient, RH) [6], and robustness to large biasing currents (Ibias) [7]. However, the electrostatic forces between the current-biased device and metallically coated probe gives rise to parasitic electric field [8], [9]. The resulting measurement of the transverse voltage (Vxy) is the superposition of the electric and magnetic field contributions, making it difficult to accurately determine

Bprobe.

We present Bprobe calibration method that eliminates the parasitic electric field with the use of frequency-modulated Kelvin probe force microscopy (FM-KPFM) feedback loop [10]. This technique, performed in ambient conditions with a 1-µm wide epitaxial graphene Hall sensor, effectively separates the electric field contributions, giving rise to Vxy signal solely due to Bprobe.

II. SAMPLE FABRICATION

The epitaxial graphene was grown by sublimation of Si and subsequent graphene formation on the Si-terminated face of

4H-SiC(0001) substrate at 2000°C and 1 bar argon gas

pressure. Details of the fabrication and structural characterization are reported elsewhere [11]. The high temperature annealing process results in a substantial number of atomic scale terraces on SiC (Fig. 1a). In the sample studied

here, the terraces are around a micron wide and do not affect the continuity of the graphene layer. The sample consists of ~95% one layer graphene (1LG) and ~5% double layer graphene (2LG); as revealed by large scale FM-KPFM mapping.

The electrodes and the Hall bars were defined by electron beam lithography in three independent steps. Oxygen plasma etching was used to pattern the Hall bars. Using this method, double Hall sensors with symmetric crosses of width ranging from 500 to 1000 nm were formed and studied at room temperature using magnetotransport and noise spectral measurements. Contact mode AFM was also used to clean the device of resist residues, as the resist is known to significantly affect the transport properties [12]. In the present study, 1-μm wide cross epitaxial graphene device was used (Fig. 1a).

M

Manuscript received October 31, 2012. Corresponding author: Olga Kazakova (e-mail: olga.kazakova@npl.co.uk).

Digital Object Identifier inserted by IEEE

Fig. 1: (a) Sample morphology largely dominated by SiC terraces. The wavy lines are atomic scale steps in the SiC substrate occurring during the high temperature graphene growth. (b) Surface potential mapping using the FM-KPFM technique with device biased at Ibias = 20 µA.

(3)

III. EXPERIMENTAL METHOD

Single pass FM-KPFM utilizes tapping mode AFM and AC/DC voltages to map the sample morphology and surface potential, respectively. The difference in potential between the probe and the current biased sample gives rise to the electrostatic force

2

2

V

dz

dC

F

, where (1)

)

2

cos(

mod mod

f

t

V

V

V

V

probe

CPD

, (2) where C and z are probe-sample capacitance and spacing,

Vprobe and Vmod are the DC and AC components of the voltage applied to the probe and VCPD is the contact potential difference [10]. The topography is measured at the mechanical resonance frequency (f0 ~300 kHz with an oscillation amplitude set point of 10-20 nm), while a low frequency AC voltage (Vmod = 5 V at fmod ~2 kHz) applied simultaneously to the electrically conductive probe. Vmod gives rise to sideband resonances of frequencies f0 ± fmod produced by oscillating electrostatic force gradient (dF/dz), which is expressed as

 

dz

dF

k

f

f

f

2

1

1

0 mod 0 , (3)

where k is the spring constant. The FM-KPFM feedback loop minimises the sidebands by applying Vprobe such that Vprobe

VCPD = 0, where VCPD is the contact potential difference, therefore eliminating the electric field between the probe and sample. Recording Vprobe pixel by pixel provides the mapping of the surface potential (Fig. 1b), which can ultimately be used to determine the work function of the sample [13].

The experimental method consisted of scanning the current biased (Ibias = 20 µA) Hall sensor with FM-KPFM technique using conductive magnetic scanning probes, while simultaneously measuring Vxy with an external Stanford Research SR830 lock-in amplifier, referenced to the mechanical oscillation of the cantilever (Fig. 2). Due to the finite potential of the biased sensor (~1 V), the grounded probe acts as a local scanning gate that couples capacitively to the sample. The FM-KPFM feedback loop accurately accounts for the surface potential, hence eliminating electrostatic forces between the probe and sensor, resulting in the measurement of only the magnetic contribution of the probe.

Using the descried method, the Vxy was mapped using two types of magnetic probes with different thickness of Co/Cr

coating: MESP and MESP-HM (Bruker) probes with coercivity ~400 Oe and the moment m ~1×10-13 and >3×10-13 emu, respectively.

IV. RESULTS AND DISCUSSIONS A. Transport Measurements

Transport and noise measurements were performed to fully characterise the Hall sensor in ambient conditions, which will be used to calibrate the stray field of the magnetic probes. The sensitivity of the sensor was determined by sweeping the DC magnetic field (B) up to ~0.55 T, while simultaneously measuring the Hall voltage (VH) for bias currents of Ibias = 10, 30 and 50 µA (Fig. 3). The Hall coefficient (RH = Vxy/IbiasB) was determined giving an average of RH = 1250 Ω/T. The finite VH offset at zero fields can be a result of misaligned voltage leads and/or non-uniform flow of the carriers due to material inhomogeneities. The conduction in epitaxial graphene occurs through electrons (n-type). The measured carrier density ne = 510

11

cm-2 and mobility μ = 1500 cm2/Vs are comparable to other published work [6], [14], [15]. The noise spectral density at Ibias = 20 µA reveals a noise floor of

Sn ~40 nV at f0 = 80 kHz, leading to a minimum detectable field of Bmin ~1.6 µT.

B. Scanning Gate Microscopy

First, we consider the case of standard scanning gate microscopy (SGM) with a metallic magnetic probe, i.e. where FM-KPFM feedback is disabled. For a finite electric field between the probe and sample, the dominating features in mapping of Vxy are peaks at the corners of the active sensing area (Fig. 4). Consider the case where the probe is gating above corner 1 and 4 (corner 2 and 3) of the sensor, the flow of electrons is diverted towards V (V+), resulting in a drop (rise) in Vxy. In essence, these features are identical to those observed in SGM experiments and are well documented [9], [16].

C. Magnetic Probe Calibration

Next, we perform SGM mapping of the device with the FM-KPFM feedback loop enabled. In the ideal case of total

Fig. 3: Room temperature field dependence of the DC Hall voltage for the 1-μm wide graphene device at Ibias = 10, 30 and 50 μA. (b) Noise spectral density at Ibias = 0 and 20 µA.

(4)

nullification of the probe-sample electric field, the response of

Vxy is a consequence of only the modulated Bprobe at f0. Then the largest Vxy is measured when the probe is at the centre of the sensing area, which is a result of its maximum coupling to the probe stray field (Fig. 5). However, we still observe small peaks at the corners of the sensor (see e.g. Fig. 5c-d). These peaks are inevitably a result of only partial nullification of the probe-sample electric field. Nevertheless, signal from Bprobe is generally unaffected and these parasitic signals have only a negligible effect on the data analysis.

Both types of magnetic probes were calibrated with forward (↓) and reverse (↑) polarities of the magnetizations, producing

a positive and negative response of Vxy, respectively. The maximum measured response of MESP probe was Vxy(↓) = 0.39 μV and Vxy(↑) = – 0.37 μV, while MESP-HM showed a larger response of Vxy(↓) = 1.08 μV and Vxy(↑) = – 0.64 μV (Table I). The resulting line profiles (see Fig. 5e) across the centre of the device, as shown in Fig. 5a by the dashed black line, demonstrate a strong signal with a Lorentzian response for data sets in Fig. 5a, 5c and 5d. However, date for MESP-HM ↑ shows signs of Vxy saturation between the positions of 0.7 and 2 µm of the line profile (Fig. 5b). Saturation for MESP-HM ↑ cannot be explained by total encompassing of the stray field as it not observed in MESP line profiles, where the probe stray field is smaller than that of MESP-HM.

D. Modelling

In order to evaluate Bprobe, micromagnetic simulations using OOMMF software were carried out for a Co coating MFM probe. The parameters used to simulate the 40 nm-thick polycrystalline cobalt layer are the following: saturation magnetization Ms = 1400 kA/m, exchange stiffness A = 3×10-11 J/m, magnetocrystalline anisotropy is negligible and a cell size of 1 nm. The geometry of the probe has been modelled following the real shape of the MESP probes with a probe radius of 30 nm. After saturating the probe along the z direction, the equilibrium magnetization distribution was simulated in a remnant state. Then, the emerging Bprobe was calculated in the surrounding volume. Fig. 6a shows the Bprobe profiles produced by the probe at different probe-sample distances and fitted to a Lorentzian function. The FWHM values of the fitted curves are shown in inset Fig. 6b. Since the measured VH values are proportional to the total Bprobe, the integral of the Bprobe has been calculated. However, since the magnetic field is not homogeneous across the area of the sensor (Across) [3], it is necessary to estimate the effective area of the probe (Aeff) and the effective stray field created by the probe (Beffect) to evaluate the VH. The effective area is calculated taking the radius = FWHM/2. The effective stray field can be calculated assuming the Bprobe spread homogeneously into the calculated Aeff. Notice the exponential decay of the calculated Beffect when the probe-sample distance increases (see Fig. 6b). In order to calculate VH produced by the probe, we use the equation VH = RH Ibias Beffect A, where RH = 1250 Ω/T, Ibias = 20 µA and A = Aeff/Across.

Notice that the Vxy data measured experimentally corresponds to an AC value induced by the oscillation of the cantilever near the sensor. These calculated values have been evaluated for the experimental parameters, i.e. with the amplitude of oscillation of Aosc = 10 and 20 nm and probe-sample distances of 10 and 20 nm.

The micromagnetic simulation corresponding to MESP probes gives Bprobe of 153 and 70 mT for probe-sample distances of 10 and 20 nm, respectively. The expected Vxy values for Aosc = 10 and 20 nm are 0.74 and 0.87 µV,

Fig. 5: (a)-(d) Transverse voltage mapping of the sensor (at Ibias = 20 µA with FM-KPFM feedback enabled) showing largely the magnetic contribution of Bprobe for MESP and MESP-HM probes with forward (↓) and reverse (↑) magnetizations. (e) Line profiles for the configurations (a)-(d) obtained along the indicated dashed line in (a). Oscillation amplitude of Aosc = 10-20 nm at f0.

Fig. 4: Transverse voltage mapping of the sensor (at Ibias = 20 µA) showing significant parasitic electric field contribution when FM-KPFM feedback loop is disabled. MESP-HM probe was used.

TABLE I

SUMMARY OF PARAMETERS FOR MFM PROBES

Probe AOSC Experimental Vxy Simulated Vxy

MESP 10/20 nm ~0.38/NA μV 0.74/0.87 μV

(5)

respectively. Moreover, the simulations corresponding to the higher moment MESP-HM probes give a Bprobe of 183 and 76 mT for probe-sample distances of 10 and 20 nm, respectively. The expected Vxy values for Aosc = 10 and 20 nm are 0.96 and 1.24 µV, respectively (Table I). The simulated responses of

Vxy are in good agreement with the experimental results. V. CONCLUSION

We successfully demonstrated calibration of Bprobe using FM-KPFM feedback loop to separate the magnetic and parasitic electric field contributions. Using novel epitaxial graphene sensors allows us to improve coupling between the magnetic probe and sensor as well as to exploit the high stability of the graphene devices to larger biasing currents, which could be attributed to the unique properties of the material, i.e. carrier energy relaxation rate (or the energy loss rate for hot carriers). This is significantly higher in graphene due to much more effective electron–lattice interactions leading to emission of acoustic phonons. The described Bprobe calibration technique is a simple first step towards calibration of the probe stray field and its gradient, at the same time the method is capable of estimating the effective probe radius. Using this technique, the experimentally determined Vxy for MESP and MESP-HM probes are in good agreement with the simulations.

ACKNOWLEDGMENT

This work was partly supported by projects Concept Graphene, IRD Graphene, MetMags and CSD2010-00024.

REFERENCES

[1] T. Kebe, “Calibration of magnetic force microscopy tips by using nanoscale current-carrying parallel wires,” Journal of Applied Physics, vol. 95, no. 3, p. 775, 2004.

[2] M. Jaafar, A. Asenjo, and M. Vazquez, “Calibration of Coercive and Stray Fields of Commercial Magnetic Force Microscope Probes,” IEEE Transactions on Nanotechnology, vol. 7, no. 3, pp. 245–250, May 2008. [3] A. Thiaville, L. Belliard, D. Majer, E. Zeldov, and J. Miltat,

“Measurement of the stray field emanating from magnetic force microscope tips by Hall effect microsensors,” Journal of Applied Physics, vol. 82, no. 7, p. 3182, 1997.

[4] H. Guillou, A. D. Kent, G. W. Stupian, and M. S. Leung, “Geometries for high spatial resolution Hall probes,” Journal of Applied Physics, vol. 93, no. 5, p. 2746, 2003.

[5] V. N. Matveev, V. I. Levashov, V. T. Volkov, O. V. Kononenko, A. V. Chernyh, M. A. Knjazev, and V. A. Tulin, “Fabrication and use of a nanoscale Hall probe for measurements of the magnetic field induced by MFM tips.,” Nanotechnology, vol. 19, no. 47, p. 475502, Nov. 2008. [6] V. Panchal, K. Cedergren, R. Yakimova, A. Tzalenchuk, S. Kubatkin,

and O. Kazakova, “Small epitaxial graphene devices for magnetosensing applications,” Journal of Applied Physics, vol. 111, no. 7, p. 07E509, 2012.

[7] J. Moser, A. Barreiro, and A. Bachtold, “Current-induced cleaning of graphene,” Applied Physics Letters, vol. 91, no. 16, p. 3, 2007. [8] M. Jaafar, O. Iglesias-Freire, L. Serrano-Ramón, M. R. Ibarra, J. M. de

Teresa, and A. Asenjo, “Distinguishing magnetic and electrostatic interactions by a Kelvin probe force microscopy-magnetic force microscopy combination,” Beilstein Journal of Nanotechnology, vol. 2, pp. 552–60, Jan. 2011.

[9] A. Baumgartner, T. Ihn, K. Ensslin, G. Papp, F. Peeters, K. Maranowski, and A. Gossard, “Classical Hall effect in scanning gate experiments,” Physical Review B, vol. 74, no. 16, pp. 1–7, Oct. 2006.

[10] U. Zerweck, C. Loppacher, T. Otto, S. Grafström, and L. Eng, “Accuracy and resolution limits of Kelvin probe force microscopy,” Physical Review B, vol. 71, no. 12, pp. 1–9, Mar. 2005.

[11] C. Virojanadara, M. Syväjarvi, R. Yakimova, L. Johansson, A. Zakharov, and T. Balasubramanian, “Homogeneous large-area graphene layer growth on 6H-SiC(0001),” Physical Review B, vol. 78, no. 24, pp. 1–6, Dec. 2008.

[12] S. Lara-Avila, K. Moth-Poulsen, R. Yakimova, T. Bjørnholm, V. Fal’ko, A. Tzalenchuk, and S. Kubatkin, “Non-volatile photochemical gating of an epitaxial graphene/polymer heterostructure,” Advanced Materials, vol. 23, no. 7, pp. 878–82, Feb. 2011.

[13] V. Panchal, T. L. Burnett, R. Pearce, K. Cedergren, R. Yakimova, A. Tzalenchuk, and O. Kazakova, “Surface potential variations in epitaxial graphene devices investigated by Electrostatic Force Spectroscopy,” 2012 12th IEEE Conference on Nanotechnology (IEEE-NANO), pp. 1–5, 2012.

[14] A. Tzalenchuk, S. Lara-Avila, A. Kalaboukhov, S. Paolillo, M. Syväjärvi, R. Yakimova, O. Kazakova, T. J. B. M. Janssen, V. Fal’ko, and S. Kubatkin, “Towards a quantum resistance standard based on epitaxial graphene,” Nature Nanotechnology, vol. 5, no. 3, pp. 186–9, Mar. 2010.

[15] C. Dimitrakopoulos, Y.-M. Lin, A. Grill, D. B. Farmer, M. Freitag, Y. Sun, S.-J. Han, Z. Chen, K. a. Jenkins, Y. Zhu, Z. Liu, T. J. McArdle, J. a. Ott, R. Wisnieff, and P. Avouris, “Wafer-scale epitaxial graphene growth on the Si-face of hexagonal SiC (0001) for high frequency transistors,” Journal of Vacuum Science & Technology B: Microelectronics and Nanometer Structures, vol. 28, no. 5, p. 985, 2010. [16] L. Folks, A. S. Troup, T. D. Boone, J. A. Katine, M. Nishioka, M. Grobis, G. J. Sullivan, A. Ikhlassi, M. Field, and B. A. Gurney, “Near-surface nanoscale InAs Hall cross sensitivity to localized magnetic and electric fields,” Journal of Physics; Condensed Matter, vol. 21, no. 25, p. 255802, Jun. 2009.

Fig. 6: Micromagnetic simulations for a Co probe of (a) stray field distribution versus the probe-sample distance. (b) Evolution of the Beffect and VH produced by the probe in the sensor with the distance. Inset: FWHM and Beffect versus the probe-sample distance.

References

Related documents

The outcome of this simulation for a constant concentration of red beads but varying concentration of green beads, Table 4.2(a), indicates that the sizing should work for both

Keywords: abetalipoproteinemia, DNA sequencing, multiplex ligation-dependent probe amplification, microsomal triglyceride transfer protein, mutation screening, paternity

Key words: Scanning probe microscopy, atomic force microscopy, spectrin, plasma- surface interactions, surface characterization and roughness, fusion, graphite, deuterium... The

Alla respondenter har haft likheter och skillnader i sitt sätt att arbeta, men slutsatsen av resultatet är att det är väldigt varierande hur lärare inom idrott och hälsa i

Även denna sponsring och produktplacering kan kopplas till indirekt marknadsföring där företag vill skapa en känsla för produkten och varumärket.. Eftersom alla undersökta

Using S&P 500 returns data, it is demonstrated that the asymmetric power autoregressive conditionally heteroskedastic volatility model has difficulty in addressing distribu-

Om den unge skulle begå ett nytt brott under tiden för verkställighet kan domstolen, istället för att påföra ett nytt straff, förlänga verkställigheten för ungdomssanktionen

Hall effect magnetic sensors have gradually gained dominance in the market of magnetic sensors during the past decades. The compatibility of Hall sensors with conventional CMOS