Preprint

This is the submitted version of a paper presented at 10th Micronano Systems Workshop (MSW), 15-16 May, 2014, Uppsala, Sweden.

Citation for the original published paper:

Andersson, M., Knaust, S., Ogden, S., Hjort, K., Bodén, R. (2014) Integrated high-pressure fluid manipulation in microfluidic systems. In:

N.B. When citing this work, cite the original published paper.

Permanent link to this version:

(2)

INTEGRATED HIGH-PRESSURE FLUID MANIPULATION IN

MICROFLUIDIC SYSTEMS

Martin Andersson, Stefan Knaust, Sam Ogden, Klas Hjort and Roger Bodén Uppsala University, Department of Engineering Sciences, Uppsala, Sweden

E-mail: roger.boden@angstrom.uu.se

Abstract

This paper presents microfluidic systems with integrated miniaturized high-pressure valves and pumps. The valves and pumps are driven by paraffin phase change material actuators. Due to its high energy density as an actuator material, paraffin is ideal for miniaturization of high pressure components. The systems are fabricated by multilayer bonding of different functional layers with stainless steel stencils as main construction material. The system integration is exemplified by two applications; the first application is a previously presented deep-sea sampler with integrated bistable high-pressure valves that handles 125 bar (12.5 MPa). In the second example we present a pump system for reduced flow variations and integrated sample injectors for high performance liquid chromatography. The combination of this multilayer fabrication method with paraffin actuation shows great potential for achieving high-pressure microfluidic systems with integrated fluid manipulation.

Introduction

Integrated microfluidic manipulation is often identified as an area that can realize the true potential of microfluidics and micro total analysis systems (µTAS). Different solutions for miniaturized pumps and valves have been presented [1-3], with various degrees of integration.

As all interfaces add both dead volume and risk of leakage, all components should ideally be situated on the same chip. For high-pressure microfluidics like high performance liquid chromatography (HPLC) and microreactors, one of the main challenges in creating such a chip is the integration of high-pressure pumps and valves [4].

In microfluidic applications, valves are often crucial components. They are generally divided into passive and active types [3], with bistable or latchable valves as a sub-group of active valves. Passive valves change states using the systems’ internal energy, whereas active valves demand external energy input for switching to and holding its non-equilibrium state. Consequently, active valves are easier to adapt to the specific needs of a system. Latchable and bistable valves combine the versatility of an active valve with low power consumption by only demanding power during the switching between states. This is important in power-restricted applications still demanding the higher performance level associated with active valves. Especially in high-pressure microfluidics, the combination of high leak pressure, low power consumption and timely switching is desirable.

The most common approach to integrated high-pressure manipulation is either electrolysis-based electrochemical (EC) pumps [5-6], or electro-osmotic flow (EOF) controlled valves [7] and pumps [8-10]. Drawbacks associated with EOF pumps are high voltages as well as the need for a certain range of pH and conductance, limiting the samples and solvents that can be used for analysis. EC pumps do not have the same limitations regarding samples and solvents. Instead, the challenges are oxidizing electrodes and current stability. Despite reports of EC pumps being able to handle pressures of 200 MPa [6], most integrated EC-based pumps reside in the region of 0.7-1.0 MPa, typically limiting the applicability, e.g. limited column length in HPLC.

For high-pressure capability of miniaturized pumps and valves, the actuator material is an important factor. To ensure large enough forces and strokes of the miniaturized and integrated actuator, an actuator material with a high energy density is needed. Phase change materials, e.g. paraffin, have the highest energy density among common actuator materials, Figure 1, for repeatable use and hence they are an attractive choice for high pressure applications [11].

Paraffin actuated pumps and valves rely on the expansion associated with the solid-liquid phase transition of paraffin. These pumps have demonstrated pressures up to 13 MPa [12-14] without affecting the flow rate and active valves can handle more than 20 MPa [15]. For bistable valves a

(3)

In the following, two different concepts of microfluidic systems with integrated high-pressure fluid manipulation are exemplified. The first is a microfluidic deep-sea sampler for a miniaturized submersible with integrated high-pressure valves from the group of Thornell [17]. The valves are bistable for lower energy consumption in energy limited applications. As the second concept we present a pump system for HPLC with integrated pumps for both the mobile phase and the sample injection.

Operating principle

The core of these paraffin actuated pumps and valves is the actuator. It consists of paraffin that is encapsulated with a rigid surrounding and a flexible membrane. When paraffin is melted by an integrated heater it expands and pushes the membrane to a deflected state. When the paraffin is let to solidify, by deactivating the heater, it contracts which returns the membrane to its original state. In this manner, the paraffin actuator membrane is used to close and open valves and perform pumping actions.

By using different geometries and by controlling the solidification of paraffin it is also possible to accomplish bi or multistable actuators that only consume energy when changing state, e.g. when opening or closing a valve.

Fabrication

The fabrication of actuators, valves, pumps and systems are made by stacking and bonding different functional layers, exemplified in Figure 2. The main layers are stainless steel stencils as structural material for actuators and fluidic channels. Additional layers can be added for different functions, e.g. flexible printed circuit boards for heaters, thermistors and integration of electronics, and glass for optical functions.

All stainless steel sheets are photochemically machined (Precision micro Ltd., UK) to define geometries, whereas the membrane was milled in a circuit board plotter (Protomat S-100, LPKF Laser & Electronics AG, Germany), and the heater layer was first wet etched (Na2(SO4)2:H2O 20g:100ml,

50°C) to define the resistive heaters from the copper cladding and subsequently etched in a reactive ion etch (15 sccm O2, 5 sccm CHF3, 150 W, 200 mTorr) to define through-holes and outer geometry.

All parts, except the covering back plate were coated with Parylene C (LabTop model 3000, Para Tech Coating Inc., USA), functioning as a bonding agent. The coated sheets were stacked and thermocompressively bonded in two steps (30 min, 200°C at 90 mbar and 30 min, 240°C at atmospheric pressure). Paraffin (CAS 8002-74-2, Sigma-Aldrich Co., USA) was casted into its designated cavities and excess paraffin was removed using a razor. Finally the back stencil was glued (Loctite 407) to the stack to seal the structure.

(4)

Applications

Microfluidic deep-sea sampler

The strong latchable valve from [16] has been modified and integrated into a microfluidic sampler by the group of Thornell [17]. The sampler was intended for a miniature submersible, enabling microbial sampling in previously unreachable underwater environments. Here, the need was a high-pressure bistable valve, managing the harsh demands of deep-sea sampling while maintaining long closing times. The sampler is shown in Figure 3. It has a sample collector formed in the top stainless steel stencil and a glass window for optical analysis of the trapped substances. Inside it incorporates an ultrasonic particle trap and bistable valves on inlet and outlet that secure the trapped sample for later analysis.

The valve managed to stay closed without power supplied at an applied pressure kept above 2.1 MPa for 19 hours, whereupon slight (50 nL/min) leakage was observed. This leakage was remedied by reactivating the valve, proving that the valve can be used for longer timeframes as long as it is regularly reactivated. The reactivation doesn’t induce additional leakage during latching. For shorter timeframes, the valve was operated up to 12.5 MPa without leakage. The valve’s burst pressure was determined to 20 MPa.

A separate valve was dedicated for measuring the capability to seal against air pressure in [18]. This valve managed to stay closed without power consumption up to 5.3 MPa of air pressure and 5.7 MPa of water pressure. The difference in pressure capability compared to the first valve may be explained by air trapped in the paraffin chamber.

Figure 2. Example from the deep see sampler of stacking of stainless steel stencils, flexible printed

circuit boards and glass, adapted from [17].

(5)

Miniaturized high performance liquid chromatography

Earlier publications [12-14] have shown the pressure capability of stainless steel paraffin-actuated micropumps. However, since all have been peristaltic pumps, a fairly high flow rate variation has been present, which induces peak broadening in HPLC and variations in reagent availability in microreactors, and consequently makes these pumps unsuitable for these purposes.

In the system presented here, we combine the operation of two integrated micropumps for solvent delivery, Figure 4, to reducing the flow rate variation significantly [19]. The pumping of individual and combined operation is seen in Figure 5.

In addition, two smaller pumps are integrated for versatile and well-controlled sample injections. The targeted application is initially miniaturized HPLC. Additionally, the miniaturized pump system can be adapted for other microfluidic high-pressure applications as well.

The multi-pump system chip for HPLC is built from stainless steel sheets, a polyimide membrane, and an integrated heater layer made from a sheet of copper-clad polyimide. The overall dimensions of the chip are 40 x 25 x 1 mm3. The designed stroke volume of the injector pumps is 60 nL. For sample

delivery, the channel between the two injector pumps can be filled by these pumps and used as a sample loop designed to 50 nL. Alternatively, one of the injector pumps can be used to inject a sample directly into the main channel, either its full volume or a variable amount.

Figure 4. Left: Photograph of the fabricated chip. Right: Top view of the system, showing the channel

system, and the twelve actuators that drives the two mobile phase pumps and two sample injection pumps.

Figure 5. Pumping performance of the system. (a) and (b): Pump 1 and 2 respectively driven

(6)

Conclusion

Integration of high-pressure fluid manipulation has been successfully shown in two different system concepts. The combination of the multilayer fabrication technique based on stainless steel stencils with paraffin actuation has great potential for realization of high-pressure microfluidic systems with integrated fluid manipulation.

References

[1]

Laser D J, Santiago J G 2004 A review of micropumps J Micromech Microeng 14(6):R35–R64

[2]

Iverson B D, Garimella S V 2008 Recent advances in microscale pumping technologies: a review and evaluation Microfluid Nanofluid 5(2) 145–74

[3]

Oh KW, Ahn CH 2006 A review of microvalves J Micromech Microeng 16(5) R13–R39

[4]

Lavrik N V, Taylor L T, Sepaniak M J 2011 Nanotechnology and chip level systems for pressure driven liquid chromatography and emerging analytical separation techniques: A review Anal Chim Acta 694 6

[5]

Xie J, Miao Y, Shih J, Tai Y C, Lee T D 2005 Microfluidic platform for liquid chromatography-tandem mass spectrometry analyses of complex peptide mixtures Anal Chem 77 6947

[6]

Fuentes H V, Wooley A T 2007 Electrically actuated, pressure-driven liquid chromatography separations in microfabricated devices Lab Chip 7 1524

[7]

Kirby B J, Shepodd T J, Hasselbrink E F 2002 Voltage-addressable on/off microvalves for high-pressure microchip separations J Chromatogr A 979 147-54

[8]

Lazar I, Trisiripisal P, Sarvaiya H A 2006 Microfluidic liquid chromatography system for proteomic applications and biomarker screening Anal Chem 78 5513

[9]

Borowsky J F, Giordano B C, Lu Q, Terray A, Collins G E 2008 Electroosmotic flow-based pump for liquid chromatography on a planar microchip Anal Chem 80 8287

[10]

Dutta D, Ramsey M 2011 A microfluidic device for performing pressure-driven separations Lab Chip 11 3081

[11]

Ogden S, Klintberg L, Thornell G, Hjort K, Bodén R 2013 Review on miniaturised paraffin phase change actuators, valves, and pumps Microfluid Nanofluidics doi:10.1007/s10404-013-1289-3

[12]

Bodén R, Hjort K, Schweitz J-Å, Simu U 2008 A metallic micropump for high-pressure microfluidics J Micromech Microeng 18 115009

[13]

Svensson S, Sharma G, Ogden S, Hjort K, Klintberg L 2010 High-pressure peristaltic membrane micropump with temperature control J Microelectromech Syst 19 1462

[14]

Bodén R, Ogden S, Hjort K 2013 Microdispenser with continuous flow and selectable target volume for microfluidic high-pressure applications J Microelectromech Syst doi:10.1109/JMEMS.2013.2279976

[15]

Sharma G, Svensson S , Ogden S, Klintberg L, Hjort K 2011 High-pressure stainless steel active membrane microvalves J Micromech Microeng 21 075010

[16]

Ogden S, Bodén R, Hjort K 2010 A latchable valve for high-pressure microfluidics J Microelectromech Syst 19(2) 396-401

[17]

Jonsson J, Ogden S , Johansson L , Hjort K, Thornell G 2012 Acoustically enriching, large-depth aquatic sampler Lab Chip 12 1619-28

[18]

Ogden S, Jonsson J, Bodén R, Thornell G, Hjort K 2012 Latchable Valve for Microfluidic Sampling from a Miniature Submersible Proc Actuator 2012 717-20

[19]

Ogden S, Knaust S, Dahlin A, Hjort K, Bodén R On-chip pump system for high-pressure microfluidic applications Proc microTAS 2013 1472-4

References

Related documents

This has been shown through characterizations of worn coatings used in continuous metal cutting where it is seen that the spinodally decomposed domains are

A cap eccentricity function of relative density was found using inverse modelling adjusted to the load–displacement data from the experiments and a new bulk modulus as

In other hand the relation between the ellipse diameter and applied force for the double armoured DC cable became better than the bundled AC cable, therefore it is a possible way

The figure shows the relation of negative streamwise vorticity (corner separation vortex and scraping vortex) and pos- itive streamwise vorticity (cavity vortex) to leakage mass

www.liu.se Hugo W ärner High T emper atur e F atigue Behaviour o f Aust enitic St ainless St eel

Luktmätningarna utfördes med luktpanelen i ett mobilt luktlaboratorium där objektsluft (luft från komprimatorn) presenterades i en exponeringshuv och jämfördes med olika

Majoriteten av tidigare forskning behandlar TPB på antingen yngre eller äldre individer, och inte ofta i forskning där yngre och äldres entreprenöriella intention studeras i

For constants that are read by all or many of the threads, like a filter kernel in image or volume convolution, the constant memory should be used since it is cached and thereby