• No results found

Count rate linearity and spectral response of the Medipix3RX chip coupled to a 300μm silicon sensor under high flux conditions

N/A
N/A
Protected

Academic year: 2022

Share "Count rate linearity and spectral response of the Medipix3RX chip coupled to a 300μm silicon sensor under high flux conditions"

Copied!
10
0
0

Loading.... (view fulltext now)

Full text

(1)

Journal of Instrumentation

OPEN ACCESS

Count rate linearity and spectral response of the Medipix3RX chip coupled to a 300μm silicon sensor under high flux conditions

To cite this article: E Frojdh et al 2014 JINST 9 C04028

View the article online for updates and enhancements.

Related content

Technical Fundamentals of Radiology and CT: Complete tomographic installations G A Cervantes

-

Comparisons between simulation and measurements taken with the Medipix3RX detector

J McGrath, J Marchal, R Plackett et al.

-

Germanium ``hexa'' detector: production and testing

M. Sarajli, D. Pennicard, S. Smoljanin et al.

-

Recent citations

New architecture for the analog front-end of Medipix4

V. Sriskaran et al -

Monte Carlo simulation of hybrid pixel detectors

D.P. Magalhaes and A. Tomal -

1.2 Mfps standalone X-ray detector for Time-Resolved Experiments P. Maj et al

-

This content was downloaded from IP address 79.136.38.171 on 07/08/2020 at 12:35

(2)

2014 JINST 9 C04028

PUBLISHED BYIOP PUBLISHING FORSISSAMEDIALAB

RECEIVED: November 28, 2013

REVISED: December 9, 2013

ACCEPTED: March 26, 2014

PUBLISHED: April 22, 2014

15th INTERNATIONAL WORKSHOP ONRADIATION IMAGING DETECTORS

23–27 JUNE 2013, PARIS, FRANCE

Count rate linearity and spectral response of the Medipix3RX chip coupled to a 300µm silicon sensor under high flux conditions

E. Frojdh,a,b R. Ballabriga,bM. Campbell,bM. Fiederle,c,d E. Hamann,c,d T. Koenig,c X. Llopart,bD. de Paiva Magalhaes,e and M. Zuberc

aMid Sweden University,

Holmgatan 10, 851 70, Sundsvall, Sweden

bCERN,

Route de Meyrin 385, Geneva, Switzerland

cInstitute of Photon Science and Synchrotrom Radiation & ANKA Synchrotron Radiation Facility Karlsruhe Institute of Technology KIT,

Hermann-von-Helmholtz-Platz 1 D-76344 Eggenstein-Leopoldshafen Germany

dFreiburger Materialforschugszentrum FMF,

Stefan-Meier-Straße 21 D-79104 Freiburg i. Br. Germany

eBrazilian Synchrotron Light Laboratory LNLS,

Caixa Postal 6192 CEP 13083-970, Campinas - SP, Brazil E-mail:erik.frojdh@cern.ch

ABSTRACT: For clinical X-ray imaging, the detector performance under high flux conditions is very important, with typical flux rates for modern CT systems reaching 109photons s−1mm−2in the direct beam. In addition, for spectral imaging a good energy resolution under these conditions is needed. This poses difficulties, since pulse pileup in the pixel electronics does not only affect the count rate, leading to a deviation from the otherwise linear behavior, but also degrades the spectral response of the detector, making k-edge subtraction and other contrast enhancement techniques less efficient. In this paper, we investigate the count rate capabilities and the energy response of the Medipix3RX chip under high flux conditions using 10 keV monochromatic photons.

KEYWORDS: X-ray detectors; Hybrid detectors

Corresponding author.

(3)

2014 JINST 9 C04028

Contents

1 Introduction 1

2 Medipix3RX 1

2.1 Gain modes 2

2.2 IKRUMcurrent 2

3 Paralyzable detector model 2

4 Experimental setup 2

5 Results 3

5.1 Low flux energy resolution 3

5.2 Dead time measurements 3

5.3 Spectral response at high flux 4

6 Conclusions 5

1 Introduction

Single photon processing detectors offer several advantages over energy integrating detectors such as being free from non photonic noise [1, 2], giving the possibility of energy weighting of pho- tons [3] and k-edge imaging [4]. However, the processing of the pulse from each photon takes a finite time. If another hit occurs during this time, it can either be lost or added to the first pulse which will then lead to an underestimation of the photon flux and a distorted energy response.

2 Medipix3RX

The Medipix3RX chip is a single photon processing hybrid pixel detector, developed in the frame- work of the Medipix3 collaboration. The pixels matrix consists of 256 × 256 pixels with a pixel size of 55 × 55 µm2. There are two analog thresholds and two counters in each pixel. The chip can be used either in single pixel mode (SPM) where each pixel works independently or in charge summing mode (CSM) where the charge from a single interaction is summed over a dynamically allocated 2x2 pixel cluster.

In addition to the intrinsic 55 µm pixel pitch the Medipix3RX can also be used with 110 µm pixel pitch, combining thresholds and counters in four pixels. This gives the possibility to use up to eight thresholds in single pixel mode and four thresholds in charge summing mode.

Details on the chip and different operating modes can be found in R. Ballabriga et al. [5]

– 1 –

(4)

2014 JINST 9 C04028

2.1 Gain modes

The Medipix3RX [5] chip offers four different gain modes, implemented by selecting the value of the preamplifier’s feedback capacitance, 7fF for the highest gain, 14 fF, 21 fF and 28 fF for the low- est gain. In this paper, the four gain modes are referred to as super high gain mode (SHGM), high gain mode (HGM), low gain mode (LGM) and super low gain mode (SLGM) . All the dead time measurements presented here were done in HGM, but some of the energy resolution measurements were performed in SHGM as well.

2.2 IKRUMcurrent

The Medipix3RX chip uses a preamplifier scheme proposed by Krummenacher [10]. The charge sensitive amplifier is based on a cascoded differential CMOS preamplifier. The preamplifier output return to zero is controlled with a bias current IKRUMwhich is sent globally to the pixels. The value of this current affects the count rate capabilities, and as a second order effect the gain of the chip.

3 Paralyzable detector model

A number of models are commonly used to describe a detector’s performance under high fluxes. [6–

9] In this paper, a simple paralyzable detector model [6], shown in equation (3.1) is used to fit the count rate data (x) and extract the dead time (µ).

f(x) = xe−µx (3.1)

4 Experimental setup

The measurements presented in this study were performed at the TOPO-TOMO beam line at the ANKA synchrotron. Monoenergetic X-rays from 6 keV to 15 keV were used and the sensor was flood illuminated. Two detectors were investigated, one with 55 µm pixel pitch and one with 110 µ m pixel pitch. Both detectors had a thickness of 300 µ m. The sensor material was silicon, which we chose to provide a clean signal in order to investigate the ASIC rather than effects in the sensor layer. To control and read out the chip a Fitpix USB Readout System [11] was used in combination with a custom software.

The photon flux was varied with a combination of Aluminium filters and in addition because the beam was not homogeneous, multiple points could be extracted from different areas of the sensor using the same measurement. The incoming photon flux was measured for the low flux case, thus in the linear range, using the Medipix3RX chip and then calculated from the change of filtration for higher fluxes.

Before the measurements the detectors were equalized using test pulses to optimize the en- ergy response but no offline alignment of the spectrum or gain correction has been done. Data analysis including the fitting of the dead times was carried out using a combination of Python and ROOT [12].

(5)

2014 JINST 9 C04028

4 6 8 10 12 14

Energy (keV) 0.0

0.2 0.4 0.6 0.8 1.0

Counts (Differentiated, Normalized)

SPM 55um FWHM: 1.37keV SPM 110um FWHM: 1.43keV CSM 55um FWHM: 2.03keV CSM 110um FWHM: 2.20keV

Figure 1. Energy response of the Medipix3RX to 10 keV monochromatic photons. The measurement was done using a threshold scan and the normalized peak is fitted with a single Gaussian.

5 Results

5.1 Low flux energy resolution

Before the dead time measurements the energy response under low flux conditions was measured.

This sets a baseline performance to compare against during the high flux measurements. As shown in figure 1, the detectors show similar energy resolution but the charge sharing is, as expected significantly lower for the 110 µm detector. In charge summing mode, charge sharing is fully suppressed for both detectors but the energy resolution is slightly degraded because the chip sums not only the charge from four pixels but also the uncorrelated noise in quadrature. It can also be seen that the Medipix3RX charge summing over an 110 × 110 µm2 area is superior to fixed 110 × 110 µm2pixels. This is due to the dynamic allocation of the summing area being more likely to capture the full charge.

The measured energy resolution for the different operating points used in the experiment is summarized in table1. The best energy resolution, a full width at half maximum, (FWHM) of 0.96 keV was measured at 12 keV, in single pixel mode using the highest gain setting, (SHGM). This is achieved directly off the chip without any offline alignment of the data.

5.2 Dead time measurements

The count rate linearity was measured for IKRUM DAC settings of 5, 25 and 100, using 10 keV photons and an energy threshold set to 7 keV. The slightly high threshold of 7keV was chosen so that the same threshold could be maintained trough the different IKRUM settings since we did not achieve noiseless operation at IKRUM 100 and a 5 keV threshold at high flux. Figure2 shows the measurement points and the fitted dead times for the 55 µm detector and in table2the fitted dead times are presented. When compared to the 110 µm detector, we see no difference in the count rate capability per pixel but due to the larger pixels it means that the 110 µm detector can handle four times less photons per unit area. In figure3the reponse in counts pixel−1s−1is presented. As expected, for all operating modes increasing IKRUMvalues lead to a better count rate linearity.

– 3 –

(6)

2014 JINST 9 C04028

Table 1. Measured energy resolutions for the various modes of operation.

Configuration FWHM (keV)

Pixel Size Gain Mode Operating Mode IKRUM 6 keV 10 keV 12 keV 15 keV

55µm HGM SPM 5 1.30 1.37 1.36 1.37

55µm HGM CSM 5 1.85 2.03 1.95 2.2

55µm HGM SPM 25 1.69 1.51 1.6 1.61

55µm HGM CSM 25 2.23 2.44 2.41 2.63

55µm HGM SPM+PZC 25 1.72 1.65 1.45 1.72

55µm HGM CSM+PZC 25 2.26 2.52 2.46 2.66

55µm SHGM SPM 5 1.12 - 0.96 -

55µm SHGM SPM 25 1.54 - 1.3 -

110µm HGM SPM 5 - 1.43 - 1.53

110µm HGM CSM 5 - 2.2 - 2.41

110µm HGM SPM 25 - 1.62 - 1.72

110µm HGM CSM 25 - 2.66 - 2.74

110µm HGM SPM+PZC 25 - 1.55 - 1.66

110µm HGM CSM+PZC 25 - 2.81 - 2.92

110µm HGM SPM 100 - 2.49 - 2.62

110µm HGM CSM 100 - 4.02 - 4.11

Table 2. Measured dead times for the different operating modes along with the point at which a dead time loss of 10% occurs

10% dead time loss (photons ·mm−2· s−1) Mode IKRUM Dead Time 55 µm Pixels 110 µm Pixels

SPM 5 0.69 µs 5.05 · 107 1.26 · 107

SPM 25 0.57 µs 6.11 · 107 1.53 · 107

SPM 100 0.40 µs 8.17 · 107 2.18 · 107

CSM 5 3.50 µs 9.95 · 106 2.49 · 106

CSM 25 2.88 µs 1.21 · 107 3.02 · 106

CSM 100 2.02 µs 1.72 · 107 4.31 · 106

5.3 Spectral response at high flux

The count rate deviation can be modeled and corrected for, but the spectral response is also affected, an effect, which in general cannot be compensated for. [13] As shown in figures4,5,6and7, the single pixel mode is, as expected, much less affected by high count rates than the charge summing mode. This is because in charge summing mode, when a hit is assigned to a pixel its neighbors are inhibited during the process. In figure8the energy response of the 55 µ and the 110 µm detector is compared at approximately the same flux. Due to larger pixels there is more pileup in the 110 µ m detector but on the other hand there is less charge sharing. The charge summing mode in the 55 µm detector shows less charge sharing than the single pixel mode in the 110 µm but there is slightly more pileup.

(7)

2014 JINST 9 C04028

107 108

Input Count Rate (counts/mm2/s)

107 108

Output Count Rate (counts/mm2/s)

Ideal response SPM Ikrum 100 Paralyzable function Dead time: 0.4µs SPM Ikrum 25 Paralyzable function Dead time: 0.57µs SPM Ikrum 5 Paralyzable function Dead time: 0.69µs CSM Ikrum 100 Paralyzable function Dead time: 2.02µs CSM Ikrum 25 Paralyzable function Dead time: 2.88µs CSM Ikrum 5 Paralyzable function Dead time: 3.5µs

Figure 2. Measured and fitted count rates for the 55 µm detector in SPM and CSM at IKRUMDAC settings of 5, 25 and 100.

104 105 106

Input Count Rate (counts/pixel/s) 104

105 106

Output Count Rate (counts/pixel/s)

Ideal response 55um, SPM Ikrum 25 Paralyzable function Dead time: 0.57µs 110um, SPM Ikrum 25 Paralyzable function Dead time: 0.66µs 55um, CSM Ikrum 25 Paralyzable function Dead time: 2.88µs 110um CSM Ikrum 25 Paralyzable function Dead time: 2.94µs

Figure 3. Comparison between the 55 µm and 110 µm detector.

6 Conclusions

Both the count rate linearity and the energy response of the Medipix3RX chip were measured for several operating points. The charge summing mode fully suppresses charge sharing in the investi- gated energy range but the count rate capabilities are about 4-5 times less than for the single pixel mode, as expected. The measurements for count rate linearity agrees with similar measurements done by T. Koenig et al. [14] using a Medipix3RX chip with a CdTe sensor. We could not find a notable difference in the count rate capabilities between the 55µm pixel sensor and the 110 µm sensor that would go beyond the factor of four due to the larger pixel areas.

– 5 –

(8)

2014 JINST 9 C04028

Figure 4. Energy response for various X-ray fluxes for the 55 µm detector in SPM.

Figure 5. Energy response for various X-ray fluxes for the 55 µm detector in CSM

Figure 6. Energy response for various X-ray fluxes for the 110 µm detector in SPM

(9)

2014 JINST 9 C04028

Figure 7. Energy response for various X-ray fluxes for the 110 µm detector in CSM

5 10 15 20 25

Energy (keV)

0.0 0.2 0.4 0.6 0.8 1.0

Counts (Differentiated, Normalized)

CSM 55µm 1.36e7 photons/mm2/s SPM 110µm 1.17e7 photons/mm2/s CSM 55µm 4.3e7 photons/mm2/s SPM 110µm 5.14e7 photons/mm2/s

Figure 8. Comparison of 55µm pixels in CSM and 110µm pixels in SPM

Acknowledgments

We would like to thank the ANKA synchrotron for the provision of beamtime and Thorsten Mueller (beamline manager) for his support.

This research has been partly supported by the Marie Curie Initial Training Network Fellow- ship of the European Community’s Seventh Framework Program under Grant Agreement PITN- GA-4 2011-289198-ARDENT

References

[1] C. Schwarz et al., Measurements with Si and GaAs pixel detectors bonded to photon counting readout chips, Nucl. Instrum. Meth. A 466 (2001) 87.

[2] B. Mikulec, Single photon detection with semiconductor pixel arrays for medical imaging applications, Ph.D. dissertation, Univ. Vienna, Vienna, Austria, June 2000.

[3] P. Shikhaliev Tilted angle CZT detector for photon counting/energy weighting x-ray and CT imaging, Phys. Med. Biol.51 (2006) 4267.

– 7 –

(10)

2014 JINST 9 C04028

[4] J. Schlomka et al., Experimental feasibility of multi-energy photon-counting k-edge imaging in pre clinical computed tomography, Phys. Med. Biol. 53 (2008) 4031.

[5] R. Ballabriga et al., The Medipix3RX: a high resolution, zero dead-time pixel detector readout chip allowing spectroscopic imaging,2013 JINST 8 C02016

[6] G.F. Knoll, Radiation detection and measurements, John Wiley and Sons, Inc., New York (2000).

[7] K. Taguchi, et al. Modeling the performance of a photon counting x-ray detector for CT: Energy response and pulse pileup effects, Med. Phys. 38 (2011) 1089.

[8] L. Wielopolski, R. Gardner, Prediction of the pulse-height spectral distortion caused by the peak pile-up effect, Nucl. Instrum. Meth. A 133 (1976) 303.

[9] K. Taguchi, et al., An analytical model of the effects of pulse pileup on the energy spectrum recorded by energy resolved photon counting x-ray detectors, , proc. of SPIE 7622 Medical Imaging 2010:

Physics of Medical Imaging, 76221C, March 18, 2010.

[10] F. Krummenacher, Pixel detectors with local intelligence: an IC designer point of view, Nucl.

Instrum. Meth.A 305 (1991) 527.

[11] V. Kraus et al., FITPix - fast interface for Timepix pixel detectors 2011 JINST 6 C01079.

[12] R. Brun, F. Rademakers,ROOT - An Object Oriented Data Analysis Framework, proceedings of AIHENP’96 Workshop, Lausanne, Sep. 1996,Nucl. Instrum. Meth.A 389 (1997) 81.

See alsohttp://root.cern.ch/.

[13] K. Rink et al., Investigating the feasibility of photon-counting K-edge imaging at high x-ray fluxes using nonlinearity corrections, Med. Phys. 40 (2013) 101908.

[14] T. Koenig et al., Charge Summing in Spectroscopic X-Ray Detectors with High-Z Sensors, IEEE Trans. Nucl. Sci., in press.

References

Related documents

The government formally announced on April 28 that it will seek a 15 percent across-the- board reduction in summer power consumption, a step back from its initial plan to seek a

Energy issues are increasingly at the centre of the Brazilian policy agenda. Blessed with abundant energy resources of all sorts, the country is currently in a

Indien, ett land med 1,2 miljarder invånare där 65 procent av befolkningen är under 30 år står inför stora utmaningar vad gäller kvaliteten på, och tillgången till,

Det finns många initiativ och aktiviteter för att främja och stärka internationellt samarbete bland forskare och studenter, de flesta på initiativ av och med budget från departementet

Den här utvecklingen, att både Kina och Indien satsar för att öka antalet kliniska pröv- ningar kan potentiellt sett bidra till att minska antalet kliniska prövningar i Sverige.. Men

Av 2012 års danska handlingsplan för Indien framgår att det finns en ambition att även ingå ett samförståndsavtal avseende högre utbildning vilket skulle främja utbildnings-,

Det är detta som Tyskland så effektivt lyckats med genom högnivåmöten där samarbeten inom forskning och innovation leder till förbättrade möjligheter för tyska företag i

Sedan dess har ett gradvis ökande intresse för området i båda länder lett till flera avtal om utbyte inom både utbildning och forskning mellan Nederländerna och Sydkorea..